Home AboutUs Products services news&events Dealers Education contactus
STATE OF ART INSTRUMENTATION FROM AN INNOVATIVE COMPANY
micro-r-survey meter - type : ur 705 Print
Technical Data

Micro-R Survey Meter Type UR 705 manufactured by NUCLEONIX SYSTEMS is primarily designed to measure low level Gamma and X-radiation. This Portable alarming Survey Meter is designed around integrally mounted 1" x 1" NaI(Tl) Scintillator to a suitable PMT, will offer an optimum performance in counting Low-Level Gamma Radiation dose rate.
This unit can measure and display dose rates in the range of 0-10000 micro-R/hr 0 -100.00 mSv/h on a dot matrix LCD Display and has auto ranging facility. Dose rates close to natural background levels can be measured accurately.
This unit is ideally suitable for Radiometric and Environmental Radiation Monitoring involving low-level radiation dose rates.
This unit employs microcontroller based design and has number of unique features. User can store upto 1000 readings and can recall back along with time stamp, for visual indication on a dot matrix LCD display. This unit has preset alarm facility which can be set throughout the range of the instrument. This unit also has a built-in RS232 port through which the stored data can be down loaded into PC. This unit has built-in preset alarm facility.

Features

  • Conforms to ANSI N42.17A performance specifications for Health Physics instrumentation (portable) & compliant.
  • Microcontroller based design
  • Compact, elegant and light weight.
  • User can store upto 1000 readings.
  • Detects Gamma and X radiation.
  • Measures dose rate from 0-10000 µR/hr / 0-100.00 µSv / hr.
  • Detector : 1" x 1" NaI (Tl) Scintillation detector probe assembly (Internal to enclosure).
  • Preset alarm facility
  • Built-in RS 232 serial port (Optional).
  • Built in audio.
  • Works on 7.5V DC dry cells.
  • RTC time stamp for each reading.
Specifications
Radiation Detected
Gamma
Detector
NaI (TI) Scintillator,1”d x 1”h coupled to a Photomultiplier Tube. Detector assembly in inside the surveymeter enclosure.
Accuracy
+/-15%
Measurement Unit
µR/h /µSv/h / cps / cpm.
Measurement Range
Dose Rate : 0-10000µR/h (or) 0-100.00 µSv/h.
Count Rate : 0-99999 cpm (or) 0-99999cps
Range Selection

Automatic

Display Indication
16x2 backlit LCD display for showing doserate/countrate apart from textual screens during configuration.
Range Selection
Senses over range & shows on display.
Data Storage
Can store upto 1000 data readings Stored data can be recalled back on to display or transmitted to PC through RS232C serial port.
RTC
Real time stamp recorded along with each reading.
Sensitivity
1 µR/hr
Preset alarm range
0-10000µR/h or
0-100µSv/h or
0-99999 cps or
0-99999 cpm
Time Constant (TC)
Low TC=2sec, Medium TC=4sec, High TC=8sec
Audio Output
Built-in piezo buzzer provides beeps for every detected pulse and alarm o/p for preset alarms..
Serial Port / Data Communication
Built-in serial port facilitates data down loading into PC. Data communication Software with connecting cable can be provided at extra cost as an additional option.
User Interface
User interface is through six control buttons START, STOP, STORE/RECALL, INC & DEC. The above buttons allow the user to program & operate the instruments in different modes.
Power
7.5 Volts, DC, size AA, MN1500 LR6 (5 X 1.5 Volts), Alkaline cells.
Fault Diagnostics
Low battery indication is provided on the display.
Type test compliance
a) Conforms to ANSI N42.17A performance specifications for Health Physics instrumentation (portable)
b) compliant
Dimensions
106W x 196L x 100Ht in mm (Approx.)
Plot No.162 A & B, Phase-II, IDA, Cherlapally, Hyderabad-500051 India Phone:+91-40-32918055, 32914548 Fax : 040 - 27262146