NaI SCINTILLATION DETECTOR(Size 3 ”x3” Well Type)-[ Type : SD 153W ]  
Technical Data

NUCLEONIX Scintillation Detector (3”x3” - Well Type) SD153W is designed to be used with Gammaray Spectrometer GR611M (Modular) or its equivalent unit.The integral assembly consists of NaI (Tl) crystal size 3”x3”, integrally coupled to a 3 inch bialkali phototube with mu-metal magnetic / light shield. The crystal and the photomultiplier tube are housed in an aluminium container.

The assembly is connected to a solid state Pre-amplifier PA311 of NUCLEONIX make. Connectors are provided on Pre-amplifier for EHT, low voltage and signal output. The complete Scintillation Detector can be housed in a Lead Shield. The pre-amplifier is designed to suit a wide range of standard Scintillation detectors. Well type detectors are recommended for Radiochemistry works.

Features

  • 3x3 NaI Integral assembly.
  • Resolution for Cs-137 better than 9%.
  • Operating Voltage 750V to 900V.
  • Recommended for use with NUCLEONIX make Gamma ray Spectrometers.
  • 3 Bialkali Phototube with mu metal magnetic / light shielding.

NaI SCINTILLATION DETECTOR WELL TYPE  3X3
Specifications
Transducer
Well type Integral assembly
Phospher
NaI (Tl), 3x3 inches crystal
Photomultiplier
3” Bialkali phototube with mu metal magnetic/light shielding.
Well Size
0.656" dia x 1.546" deep
Operating Voltage
750 - 900V
Resolution
Better than 9% with Cs-137
Pre-amplifier
Built-in
Gain (Approx)
25
Noise
(rms referred to input) Less than 50 micro volts
Output
Positive tail pulse
Output Impedance
90 Ohms
Power Requirement : (Typical)
-12V @12mA