Nal Scintillation Detector (Size : 3″x3″ Flat Type)* [ Type No. : SD153F ]
Technical Data
NUCLEONIX Scintillation Detector (3″x3″ – Flat Type) SD153F is designed to be used with Gamma Ray Spectrometer GR611M (Modular) or its equivalent unit. The integral assembly consists of NaI (Tl) crystal size 3”x3”, integrally coupled to a 3″ bialkali phototube with mu-metal magnetic / light shield. The crystal and the photomultiplier tube are housed in an aluminium container.
The assembly is connected to a solid state Pre-amplifier PA311 of NUCLEONIX make. Connectors are provided on Pre-amplifier for EHT, low voltage and signal output. The complete Scintillation Detector can be housed in a Lead Shield. The pre-amplifier is designed to suit a wide range of standard Scintillation detectors.
Features
Specifications
Transducer | Well type Integral assembly |
---|---|
Phospher | NaI (Tl), 3×3 inches crystal |
Photomultiplier | 3″ Bialkali phototube with mu metal magnetic/light shielding. | Operating Voltage | 700 – 900V |
Resolution | Better than 9.5% with Cs-137 |
Pre-amplifier | Built-in |
Gain (Approx) | 25 |
Noise | (rms referred to input) Less than 50 micro volts |
Output | Positive tail pulse |
Output Impedance | 90 Ohms |
Power Requirement (Typical) | -12V @12mA |
Application This is a flat detector used along with GR611M / GR612 – both for teaching & research purposes by R & D labs & universities.