#Radiation Survey Meter*

#Radiation Survey Meter* [ TYPE : RM 701N]

Technical Data
manual

This product is designed around a Micro-controller Chip. It is provided with alpha Numeric LCD display for indicating the doserate in digits, cumulative dose in digits. It covers wide range from 0 to 10R/hr in two range with auto ranging. This unit has a facility for storing the data into built-in E2PROM along with real time. The stored data can be recalled on to the display or transmitted to PC through USB port.

Features
Specifications
Detector GM counter Type GM 130 with energy compensated filter.
Radiation detected X & Gamma, from 60 Kev to 1.3 MeV.
Measurement Unit mR/h/µSv/h/cps/cpm.
Measuring Range 0 – 1000.0 mR/hr (Low Range)
1 R/hr – 10 R/hr (High Range)
0 – 10,000 µSv/hr (Low Range)
10 mSv/hr – 100 mSv/hr (High Range)
Preset alarm range 0-10000.0mR/r (or) 0-100000µSv/h.
Energy Response Within +/- 20% of true response from 60 KeV to 1.25 Mev.
Operation Ranges 0-10R/hr.
Over range Unit shows over range above 10R/hr to 1000R/hr.
Range Changeover Automatic
Accuracy +/- 15% with Cs-137.
Display 16X2 alpha numeric LCD display; Digital display of dose rate and units in the first line, dose display in the second line.
Response Time < 5 sec.
Serial port / data communication Built- in Micro USB port facilitates data downloading into PC. Data communication software with connecting cable can be provided at extra cost.
Audio Built-in piezo buzzer provides beeps for every detected pulse and alarm O/P for preset alarms.
Data Storage Can store upto 1000 readings. Stored data can be recalled back on to display or transmitted to PC through micro USB port.
User Interface A part from powering unit ON/OFF, STORE, INC ,DEC & ON/OFF are the seven control above buttons allow the user to program and operate the instruments in appropriate manner.
Real Time Clock Real time stamp recorded with each reading.
Fault Diagnostics Low battery indication hv fault & detector failure are provided on the display.
Type test compliance a)IP rating is better than IP56 for unit.
b)EMI/EMC compliance as perIEC61000-4-3.
c)Radiation Performance Complies toIEC61584:2001.
Dimensions 196H x 106W x 50D mm (approx).
Weight 1Kg (approx).

Application : Serves as a general purpose Radiation Survey Meter to measure Gamma Dose Rates in and around Nuclear Facilities & including Nuclear Power Plants, Environmental Survey Labs.