NAI SCINTILLATION DETECTOR (WELL TYPE) BASED GAMMA RAY SPECTROSCOPY SYSTEM WITH 8K MCA

NAI SCINTILLATION DETECTOR (WELL TYPE) BASED GAMMA RAY SPECTROSCOPY SYSTEM WITH 8K MCA -

Technical Data
manual

Essentially this system consists of the following constituent units (1) a 2”x2” well type NaI integral assembly detector with lead shielding arrangement (2) High voltage unit (3) Shaping amplifier, (4) Instrumentation bin with power supply, (5) 1K/4K/8K MCA (USB based) (6) Personel computer with data acquisition & processing software.

Features
Specifications
MINIBIN AND POWER SUPPLY MB 403:

Mini bin : Accommodates SIX / EIGHT single bit modules or combination of multiple widths with Amphenol connectors. Minibin is primarily designed with the objective of conserving bench space and to achieve significant saving in cost of the Minibin based systems. Bussed wiring is provided to the power connectors to distribute +/- 12V and +/- 24V. A control panel with ON/OFF switch, low voltage test sockets is provided on the right extreme side of the bin.

Minibin Dimensions : 11.75″width X 11.00 depth (upto connectors) X 8.75″ height.

Power supply : This is either two and half bit module or a compact box type enclosure fitted at the back of this bin, which generates highly regulated D.C voltages.

Input : (230V + 10%) a.c, 50Hz.

D.C Output : +12V @ 1A, -12V @ 1A, +24V @ 0.5A, -24V @ 0.5A 48 watts maximum.

Noise & Ripple : Less than 3 mv

Stability : +/- 0.5% after a 24 hr warmup at constant line, load & ambient temp.

HIGH VOLTAGE UNIT (HV 501) :

a. Output voltage variable continuously from 0V to 1500 volts

b. Output current (max) 1mA

c. Load & Line regulations : Better than 0.005% of full scale

d. Indefinite over load & short circuit protections and self recovery

e. Output ripple less than 20mv.

f. Dimensions : Single / Two bit module

LINEAR AMPLIFIER (LA 520) :

a. Input Polarity : Positive or Negative

b. Total gain (Typical) : 1000 (approx)

c. Output (Unipolar) : 0 to 8V

d. Max. output (Unipolar) : 12V

e. Dimensions : Two bit module

LEAD castle :

This Lead Shield is designed to shield 2”x2” NaI detector Scintillation Detectors of NUCLEONIX make. It is built-up of interlocking rings with bottom and top plates. The bottom ring is provided with a small opening so that the cables from the Scintillation Detector Pre-amplifier base could be taken out for connecting to the Gamma ray spectrometer counting system.

The inside of the lead shield is lined with Aluminium to minimise scattering.

Thickness 40mm, accommodate 2” scintillation detector including sample.

Well type scintillation detector (SD151W):

a. Crystal Sizes : 2” x 2”

b. Well Size : 0.656” dia x 1.546” deep

c. Resulation : Better than 8.5 %

d. Noise (RMS. referred to input) : Less tha 50mV

e. Operating Voltage : 700 to 900V

f. Out put : Positive Tail Pulse

g. Output impedance : 90 Ohms

h. Power Requirement (Typical) : -12V @ 12m

Multi-Channel Analyzer (8k MCA) with processing software :

Multi-Channel Analyzer (MCA) is an important part of nuclear spectroscopy system. The major requirement of MCA is for nuclear pulse height analysis in energy spectroscopy. The USB-MCA presented here, incorporates state of art technologies like FPGA,USB bus interface and precision analog electronics to meet the stringent system requirements in nuclear pulse spectroscopy. The resolution supported by the USB-MCA ranges from 256 channels to 8K channels selectable via software, making it suitable for all spectroscopy applications from low resolution (e.g. NaI-PMT) to gigh resolution (e.g.HP-Ge) systems.

The USB bus interface of the MCA provides an excellent connectivity with most of the new PCs and lap-top computers. The PHAST application software provided with the USB-MCA, seamlessly integrates with the hardware, featuring a range of standard functions required for analysis and acquisition.

SPECIFICATIONS :
Hardware features:
Software features:

Personal computer system with printer : Any standard pentium IV computer configuration with printer is adequate to run MCA software.

NaI Scintillation Detector (well type) based gamma ray spectroscopy system with 8k mca, installed in QC lab of a medical cyclotron facility. System is used to test isotope purity & energy of the Radio-nuclides such as FDG / F-18 or Tc-99, I-131etc., produced.