Dual Scaler Timer [ TYPE : CT 542A ]
Technical Data
Dual Counter Timer Type CT542A manufactured by NUCLEONIX is designed around a microcontroller chip and has 12 x 2 LCD dotmatrix display.
This unit can be used in three modes PRESET TIME, CPS & CPM modes. It has on board memory to acquire readings upto 1000. The stored data readings can be recalled onto display or can be printed onto a printer. It can count two inputs simultaneously.
This unit has a built-in parallel port (centronics) and a serial port RS 232 for connecting to a PC.
It is built as a two bit module. It can count both positive and negative pulses in the range of 100 mV to 10V and upto a maximum frequency of 1MHz. LCD display indicates preset and elapsed time and counts of any selected channel.
Programmability of ONE, TWO & FOUR iterations for any particular counting.
Features
Specifications
COUNTER / RATEMETER
Count Input(s) IN1 & IN2 | 100mV to 10V, unipolar or positive bipolar semi-gaussian pulse |
---|---|
Pulse Width | 0.5 micro sec (min) |
Polarity | Positive or Negative |
Input Impedance | 5.0 K ohms |
Input Counts Capacity | 999999 counts |
Input Frequency (max) | 1 MHz |
Pulse Height Discrimination | 100mV – 10V by a preset provided on front panel |
Modes of Data Acquisition | a. Counts for a preset time, b. CPS, c. CPM |
Counts Indication | 16 x 2 dotmatrix LCD display |
Timer
Preset Time Setting | Programmable through tactile switch control buttons |
---|---|
Control Buttons | START, STOP, PROG, STORE, INC, DEC |
Preset Time/Elapsed Time Indication | On 16 x 2 LCD Dot matrix display |
Preset Time Range | 1 to 9999 |
Printer Port | Built-in |
Serial Port | RS 232C built-in |
Additional Options (at extra cost) | a. Data communication Software for down loading of data can be given at extra cost. b. Printer. |
Module connector | Amphenol Connector Type : 26-159-24P-H (24 Pin Type) or NIM Standard as per AEC specifications TID 20893 (Rev) Type : AMP 204186-5 |