Nal Scintillation Detector (Size : 3″x3″ Well Type) * [ Type No. : SD153W ]
Technical Data
NUCLEONIX Scintillation Detector (3″x3″ – Well Type) SD153W is designed to be used with Gamma ray Spectrometer GR611M (Modular) or its equivalent unit. The integral assembly consists of NaI (Tl) crystal size 3″x3″, integrally coupled to a 3″ bialkali phototube with mu-metal magnetic / light shield. The crystal and the photomultiplier tube are housed in an aluminium container.
The assembly is connected to a solid state Pre-amplifier PA311 of NUCLEONIX make. Connectors are provided on Pre-amplifier for EHT, low voltage and signal output. The complete Scintillation Detector can be housed in a Lead Shield. The pre-amplifier is designed to suit a wide range of standard Scintillation detectors. Well type detectors are recommended for Radiochemistry works.
Features
Specifications
Transducer | Well type Integral assembly |
---|---|
Phospher | NaI (Tl), 3×3 inches crystal |
Well Size | 0.656” (dia) X 2.06” depth. |
Photomultiplier | 3″ Bialkali phototube with mu metal magnetic/light shielding. | Operating Voltage | 700 – 900V |
Resolution | Better than 9% with Cs-137 |
Pre-amplifier | Built-in |
Gain (Approx) | 25 |
Noise | (rms referred to input) Less than 50 micro volts |
Output | Positive tail pulse |
Output Impedance | 90 Ohms |
Power Requirement (Typical) | -12V @12mA |
Application This is a well type detector recommended to study Radiochemical samples in liquid form. Recommended for chemistry depts by R & D labs for radio-chemistry work, along with GR611M / GR612.