Radiation counting system with accssories for Beta sample counting (Copy)
Technical Data
Radiation Counting system, type RC605A manufactured by NUCLEONIX is a versatile state of art integral counting system designed around eight bit microcontroller chip for using with a variety of detector probes such as Alpha / Beta / Gamma scintillator detector probe or End window G.M detector probe. The system is suitable for counting Alpha, Beta and Gamma sample counting with appropriate detector probes. This system essentially works as a Beta sample counting system with any one of the probes GM125/SBP-2 connected to it.
Radiation Counting System essentially has a processor card and other electronic circuits to generate continuously variable HV upto 1500V to be applied to scintillation detector Probes ( ?, ? or ? ) or End window G.M. Tube, amplify the detector output and convert them to digital pulses for counting and displaying the recorded counts for a preset time.
System facilitates counting of samples either on planchets or filter paper.
Applications : This system will find applications for counting of air activity, wipe, environmental samples, including air, water (river, lake, pond, ground & sea waters).
System can be used by testing labs, Environmental survey labs at Nuclear Plants, in normal or in a Nuclear disaster scenario.
Features
Specifications
P.M. Input (From alpha, beta, gamma scintillation detector probe) |
(a) Polarity : Negative (b) Amplitude : -100 mV (min) |
---|---|
G.M. Input (From G.M.Counter) : | (a) Polarity : Negative (b) Amplitude : -500 mV (min) (c) Built-in load resistor : 4.7 or 3.3M Ohms |
HV Output : | HV (0-1500V) @1mA continuously variable through front panel keypad in steps of 1 volt, ripple less than 20mV, line & load regulation better than 0.05%. |
HV indication : | On LCD dotmatrix provided. |
Display : | 20 x 2 LCD dotmatrix display has been provided to indicate data counts, Elapsed Time and HV. |
Counts Capacity | 999999 counts |
Preset time | 1min to 24 hrs (HH : MM) format |
Preset cycles / Iterations : | 1 to 10 |
Command Buttons: | START, STOP, PROG, STORE, INC & DEC command buttons have been provided on the front panel key pad. |
Paralysis Time : | A choice of three paralysis times 250, 350 and 550 micro sec plus OFF position selected through PROG key. |
Programmability : | Includes selection of Preset Time, Storing / Recalling of data, starting and stopping of acquisition, label assignment for data counts BG (Background), ST (Standard) & SP (sample) etc,. |
RTC : | Built in RTC provides real time clock information which is stamped in the activity report when printed. Built in Real time clock facillitates the user to generate sample analysis reports with RTC stamping. Both date month & time in hrs and minutes are printed. |
Scintillation detector probe socket : | This is a UHF socket for connecting to a, b or g scintillation Probe. |
G.M. Socket : | UHF connector for connecting to G.M. Detector |
Printer Port : | Built-in centronics port facilitates connection to a printer for direct data printing selectively. |
Serial Port : | Built-in serial port facilitates data down loading into PC through RS232/RS485 port (MODBUS COMPLIANT) |
Data Communication Software : | (Optional at extra cost) Can be provided for serial transfer of data readings into PC. |
PCB Edge Connector : | EURO Type. |
Inter PCB Connections : | Through Mother Board. |
Power : | Unit is powered works on 230V, AC, 50Hz through power / adapter which delivers 12V input to unit. |
Operating Temperature: | 0 to 500C |
Relative Humidity : | Upto 90% |
nstrument will meet all requirements applicable to : |
(a) Manufactured confirming to ANSI N 42.17. (b)Complies to IS -9000 part III & V, for climatic test. (c) Complies to Interference test as per IEC61000 or equ. |
Mechanical Dimensions : | 250mm(W) X135mm(H)X325mm(D) Approx. |
TYPICAL ALPHA SAMPLE COUNTING REPORT
BGD CPM : 0000
CPM OF STD : 00092
DPM OF STD : 00265
EFF. OF STD : 034.7
*FLOW RATE : 01.00
PTIME (BG):0300
PTIME (ST):0300
PTIME (SP) : 0300
SL.NO | LABEL | RTC | COUNTS | P.TIME | VOL( ) | Bq/VOL | iter |
---|---|---|---|---|---|---|---|
0001 | BG | 11:43 06-09 | 0000CPM | 000180 | – | – | – |
0002 | STD | 11:52 06/09 | 000160CPM | 000060 | 0010dps | 26.56% Eff | – |
0003 | SP1 | 11:47 06/09 | 000196 | 000060 | 0000 ml- | 0000.00Bq/ml– | 01 |
0004 | SP2 | 11:48 06/09 | 000180 | 000060 | 0001ml– | 0009.03Bq/ml– | 01 |
0005 | SP3 | 11:50 06-09 | 000187 | 000060 | 0001ml– | 0000.93Bq/ml– | 01 |
Applications:This system can be used for counting ?, ? or ? samples on a 25mm dia planchet or 47/50mm dia filter paper obtained from air samplers, or continuous air monitors in a Nuclear facility. System can also used for wipe sample counting in nuclear counting lab of a Nuclear power plant or similar facility. Also this system can be used in a University for teaching lab experiments in a physics department.