Spectroscopy Amplifier

Spectroscopy Amplifier [ TYPE : SA 524 ]

Technical Data
manual

The input to SA524 can be either positive or negative signal from a detector preamplifier. The output pulses, 0 to 10V for unipolar pulse and ±10V for bipolar pulse, provided on both front and rear panel, are suitable for use with single channel and multi channel pulse height analyzers. When long connecting cables are used between detector preamplifier output and the amplifier input, noise induced in the cable from ground loop currents can reduce the signal to noise ratio thereby causing spectrum degradation.

This may be minimized by using amplifier in the differential input mode. Six different shaping time constants are provided for optimizing resolution and count rate performance. To minimize baseline shift due to drift as well as count rate dependent shift at medium and high frequencies a gated baseline restorer (BLR) has been incorporated. Automatic noise threshold stage derives a threshold voltage proportional to the noise level and ensures that BLR gating works just above noise level. The pile up reject output logic pulses may be fed to an associated multi-channel analyzer that can be used to suppress spectrum distortion caused due to piling up of pulses on each other at high counting rates.

Features
Specifications
A. PERFORMANCE
Gain Range Continuously variable from X4 to X1500.
Pulse Shaping Quasi-gaussian and Quasi-triangular.
Shaping time 0.5, 1, 2, 3, 6 and 10 micro sec.
Input Noise 5 mv r.m.s with 3 micro sec. shaping time
Overload Recovers to within 2% of baseline in 15x shaping time from x200 overload.
Integral Non-Linearity < 0.05% from 0 to 10V.
Crossover Walk Bipolar zero cross over walk is < ± 3 ns in 50:1 dynamic range.
B. CONTROLS
FINE GAIN Front panel 10 turns precision potentiometer provides a continuously adjustable, gain factor from 0.5 to 1.5.
COARSE GAIN Front panel six-position switch selects gain factors of X20, X50, X100, X200,X500 and X1000
PZ Screwdriver (Knob) adjustment of the PZ cancellation using 20-turn potentiometer on the front panel.
POS/NEG Front panel toggle switch for selecting either positive or negative input signals.
ATN A front panel toggle switch selects an input attenuation factor of X1 or X2.5
SHAPING Front panel six position switch for selecting shaping times of 0.5, 1, 2, 3, 6 and 10 micro sec.
TRI / GAUSS Front panel toggle switch for selecting quasi-gaussian or quasi-triangular unipolar output shape.
BAL Screwdriver adjustment to match the gains of normal and differential reference inputs for maximum common mode noise rejection in DIFF mode using 20 turn potentiometer on the front panel .
LIM A push-button switch on the front panel to prevent oscilloscope input from overloading and thus enabling observation of the baseline in sensitive ranges of the scope.

Note:Manufactured by NUCLEONIX SYSTEMS based on Technology from Bhabha Atomic Research Center, Mumbai.

C. INPUTS
IN BNC connectors on front panel & rear panel accepts 0 to 10V of either polarity.
D. OUTPUTS
UNI Unipolar output on front panel BNC, full-scale linear range 0 to +10V.
BI Bipolar output on front panel BNC, linear range of ±10V.
PUR Inhibit output, rear panel BNC provides a + 5V logic signal when pulse pile up occurs. This is used as an anticoincidence input for ADC.
BUSY Deadtime output, rear panel BNC provides +5V logic signal for the duration that the input pulses exceed threshold. This is used for dead time corrections with associated ADC.
E. POWER CONNECTORS
NIM standard, as per AEC specifications TID 20893 (Rev) Type AMP 204186-5 or Amphenol connector type 26-159-24P-H (24 pin type) (+24V, -24V, +12V, -12V and ground)
PREAMP POWER Rear panel power connector (9 Pin D-Type) provides preamp power (+24V, -24V, +12V, -12V and ground) to the associated preamplifier.