Detector : GM counter with energy compensated filter with external pole of 1.2meter length
Radiation detected X & Gamma, from 60 Kev to 2 MeV
Energy Response : Within +/- 20% of true response from 60 KeV to 1.25 Mev.
Operation Ranges CPS, CPM, Preset Time & Doserate Mode
Preset time range 1 sec to 9999 sec
Doserate (auto ranging) Option A : with Internal Detector
Option B : with 0 – 10 R/hr
Option C : with 0 – 100R/hr
Doserate TC Low (2 sec),
Medium (4 sec),
High (8 sec)
Over range Unit shows over range above 10R/hr to 1000R/hr
Range Changeover Automatic
Accuracy +/- 15% with Co-60.
Display : 16X2 alpha numeric LCD display; Digital display of dose rate and units in the first line, Bar graph or dose display in the second line.
Time Constant Choice of three time constants LOW, MED & HI for Doserate mode.
Serial Port (optional) Built-in RS 232 serial port for PC linking, along with data communication software cab
Aural Built-in buzzer provided
Data Storage : Storage of readings in built-in EEPROM. Stored Data can be recalled on to the display or transmitted to PC
Control Buttons START, STOP, PROG, STORE, INC & DEC are the six control buttons provided on the instrument. The above buttons allow the user to program & operate the instruments in appropriate manner.
Real Time Clock : On board real time clock with battery backup keeps real time information. On selecting data storage, current real time is saved along with doserate
Power 6V DC size AA MN1500 LR6 (4×1.5) alkaline cells
Type test compliance a) Conforms to ANSI N42.17A performance specifications for Health Physics instrumentation (portable) b) compliant
Dimensions 196H x 106W x 50D mm (approx)
RADIATION DOSERATE MEASUREMENT WITH RADIATION SURVEY METER (RM703)
USING EXTERNAL TELEPOLE DETECTOR ARRANGEMENT